RAMDHANI, ABDURRAHMAN (2020) PENENTUAN NILAI KETIDAKPASTIAN HASIL KALIBRASI DAN IMPLEMENTASINYA PADA ALAT FURNACE. S1-Sarjana thesis, Universitas Mercu Buana Jatisampurna.
|
Text
1. Halaman Judul.pdf Download (188kB) | Preview |
|
|
Text
2. Lembar Pernyataan.pdf Download (196kB) | Preview |
|
|
Text
3. Lembar Pengesahan.pdf Download (140kB) | Preview |
|
|
Text
4. Abstrak.pdf Download (104kB) | Preview |
|
|
Text
5. Abstract.pdf Download (104kB) | Preview |
|
|
Text
6. Kata Pengantar.pdf Download (161kB) | Preview |
|
|
Text
7. Daftar Isi.pdf Download (215kB) | Preview |
|
|
Text
8. Daftar Tabel.pdf Download (107kB) | Preview |
|
|
Text
9. Daftar Gambar.pdf Download (103kB) | Preview |
|
|
Text
10. Daftar Lampiran.pdf Download (103kB) | Preview |
|
![]() |
Text
11. BAB 1.pdf Restricted to Registered users only Download (162kB) |
|
![]() |
Text
12. BAB 2.pdf Restricted to Registered users only Download (393kB) |
|
![]() |
Text
13. BAB 3.pdf Restricted to Registered users only Download (288kB) |
|
![]() |
Text
14. BAB 4.pdf Restricted to Registered users only Download (272kB) |
|
![]() |
Text
15. BAB 5.pdf Restricted to Registered users only Download (291kB) |
|
![]() |
Text
16. BAB 6.pdf Restricted to Registered users only Download (234kB) |
|
![]() |
Text
17. Daftar Pustaka.pdf Restricted to Registered users only Download (162kB) |
|
![]() |
Text
18. Lampiran.pdf Restricted to Registered users only Download (234kB) |
Abstract
ABSTRAK Alat Furnace merupakan tungku panas berkapasitas 14L dengan pencapaian maksimum suhu 1200°C yang berfungsi dalam proses pengujian kadar abu di bahan karet dan pengujian lainnya. Proses pengujian dilakukan pada suhu 550°C ±10°C yang telah diatur dalam SNI/ISO 247 Penentuan Kadar Abu yang dipersyaratkan. Untuk menjaga performance suhu, maka perlu dilakukan control secara kontinu dan periodik agar penyimpangan suhu furnace dapat diketahui, selain perawatan mesin perlu juga dilakukan verifikasi terhadap suhu sebenarnya yang dinamakan aktifitas kalibrasi. Kalibrasi adalah suatu kegiatan dengan cara membandingkan suhu langsung dengan alat termokopel dan indikator suhu termometer digital. Kalibrasi diukur pada suhu yang digunakan yaitu 550°C, kalibrasi dilakukan dengan memperhatikan beberapa faktor diantaranya hasil kalibrasi alat, faktor lingkungan dan nilai resolusi furnace. Hasil dari kalibrasi furnace diperoleh nilai yang disebut koreksi dan ketidakpastian pengukuran, adalah nilai penyimpangan pembacaan alat dari nilai sebenarnya dan rentang suhu. Dari hasil kalibrasi alat diperoleh nilai koreksi = - 5.0°C dan ketidakpastian = ±1.2°C, artinya nilai aktual suhu furnace adalah = 550 + (-5.0) ±1.2 . Artinya bila nilai pembacaan indikator furnace sebesar 550°C, maka nilai sebenarnya adalah (543.8 ~ 546.2)°C. Jadi dengan melaksanakan kalibrasi furnace, bisa diketahui nilai aktual suhu furnace, sehingga bisa diambil tindakan terhadap nilai penyimpangan yang diketahui. Kata Kunci : Kalibrasi, Furnace, Ketidakpastian, Ketertelusuran, Kualitas, Statistik ABSTRACT Furnace is a heat furnace with a capacity of 14L with a maximum temperature of 1200 ° C that functions in the process of testing the ash content in rubber materials and other tests. The testing process is carried out at a temperature of 550 ° C ± 10 ° C which has been regulated in SNI / ISO 247 Determination of Ash Levels required. To maintain the temperature performance, it is necessary to control continuously and periodically so that furnace temperature deviations can be known, in addition to engine maintenance it is also necessary to verify the actual temperature called calibration activity. Calibration is an activity by comparing temperature directly with thermocouple devices and digital thermometer temperature indicators. Calibration is measured at the temperature used is 550 ° C, calibration is done by taking into account several factors including the results of the calibration tool, environmental factors and the value of the furnace resolution. The results of the furnace calibration obtained a value called correction and measurement uncertainty, is the value of the deviation reading from the actual value and temperature range. From the results of the calibration tool obtained correction value = -5.0 ° C and uncertainty = ± 1.2 ° C, meaning that the actual value of the furnace temperature is = 550 + (-5.0) ± 1.2. This means that if the reading value of the furnace indicator is 550 ° C, the actual value is (543.8 ~ 546.2) ° C. So by carrying out furnace calibration, it can be known the actual value of the furnace temperature, so that action can be taken on the known deviation values. Keywords : Calibration, Furnace, Uncertainty, Traceability, Quality, Statistic
Item Type: | Thesis (S1-Sarjana) |
---|---|
Call Number CD: | FT/IND 19 045 |
NIM: | 41614320034 |
Uncontrolled Keywords: | Kalibrasi, Furnace, Ketidakpastian, Ketertelusuran, Kualitas, Statistik |
Subjects: | 600 Technology/Teknologi > 670 Manufacturing/Manufaktur, Pabrik-pabrik 600 Technology/Teknologi > 680 Manufacture For Specific Uses/Industri Pembuatan produk untuk penggunaan tertentu |
Divisions: | Fakultas Teknik > Teknik Industri |
Depositing User: | Nasruddin Mansyur S.Hum |
Date Deposited: | 07 Nov 2020 05:41 |
Last Modified: | 07 Nov 2020 05:41 |
URI: | http://repository.mercubuana.ac.id/id/eprint/58014 |