PENGEMBANGAN SENSOR pH MENGGUNAKAN ELEKTRODA LIMBAH BATERAI DAN CAMPURAN POLIPIROL ARABIC GUM

AFFICA, SAIG NALIH (2023) PENGEMBANGAN SENSOR pH MENGGUNAKAN ELEKTRODA LIMBAH BATERAI DAN CAMPURAN POLIPIROL ARABIC GUM. S1 thesis, Universitas Mercu Buana Jakarta.

[img]
Preview
Text (HAL COVER)
01 COVER.pdf

Download (605kB) | Preview
[img]
Preview
Text (ABSTRAK)
02 ABSTRAK.pdf

Download (149kB) | Preview
[img] Text (BAB I)
03 BAB 1.pdf
Restricted to Registered users only

Download (49kB)
[img] Text (BAB II)
04 BAB 2.pdf
Restricted to Registered users only

Download (520kB)
[img] Text (BAB III)
05 BAB 3.pdf
Restricted to Registered users only

Download (611kB)
[img] Text (BAB IV)
06 BAB 4.pdf
Restricted to Registered users only

Download (1MB)
[img] Text (BAB V)
07 BAB 5.pdf
Restricted to Registered users only

Download (146kB)
[img] Text (DAFTAR PUSTAKA)
08 DAFTAR PUSTAKA.pdf
Restricted to Registered users only

Download (98kB)
[img] Text (LAMPIRAN)
09 LAMPIRAN.pdf
Restricted to Registered users only

Download (132kB)

Abstract

This research was carried out to replace a pH meter, which generally has a glass electrode that functions as a measure of the amount of ions in a sample. In general, the nature of glass that is prone to breaking makes the sensor prone to leaks, this time Ion selective electrode (ISE) based on polypyrrole and battery rod electrodes. by carrying out several treatment processes and conducting tests with potentiometric devices, in the polypyrrole coating process using the Electrodeposition method with voltage variations of 1.5 Volts, 3 Volts, 4.5 Volts, 6 Volts, 7.5 Volts, and 9 Volts Obtained optimal value results for variations in voltage at 3 Volts using the electrodeposition method with a coating time of 30 seconds, in the pH response test using the potentiometry method the slope value was 53.657 mV/decade in the pH response test using pH buffers 4, 7 and 10 where these values are in accordance with IUPAC standards. variations in the addition of Arabic gum at 0.5%, 1% and 2% to ensure the addition of what percentage of Arabic gum produces the ideal nerst value. In the selectivity test, the results obtained for interfering ions using Ba2+ with a value of

Item Type: Thesis (S1)
Call Number CD: FT/MSN. 23 057
NIM/NIDN Creators: 41318010024
Uncontrolled Keywords: Sensor pH, Elektrode limbah batang baterai bekas, Potensiometri, Polypyrrole Arabic gum
Subjects: 600 Technology/Teknologi > 620 Engineering and Applied Operations/Ilmu Teknik dan operasi Terapan
600 Technology/Teknologi > 620 Engineering and Applied Operations/Ilmu Teknik dan operasi Terapan > 621 Applied Physics/Fisika terapan
600 Technology/Teknologi > 620 Engineering and Applied Operations/Ilmu Teknik dan operasi Terapan > 621 Applied Physics/Fisika terapan > 621.8 Machine Engineering, Machinery/Teknik Mesin
Divisions: Fakultas Teknik > Teknik Mesin
Depositing User: ADELINA HASNA SETIAWATI
Date Deposited: 05 May 2023 08:02
Last Modified: 05 May 2023 08:02
URI: http://repository.mercubuana.ac.id/id/eprint/76750

Actions (login required)