PENGEMBANGAN SENSOR pH (H+₋ SELECTIVE ELECTRODE) BERBASISKAN POLYPYRROLE DAN ELEKTRODA KAIN/KARBON

PANGESTU, DIMAS (2023) PENGEMBANGAN SENSOR pH (H+₋ SELECTIVE ELECTRODE) BERBASISKAN POLYPYRROLE DAN ELEKTRODA KAIN/KARBON. S1 thesis, Universitas Mercu Buana Jakarta.

[img]
Preview
Text (HAL COVER)
01 COVER.pdf

Download (560kB) | Preview
[img]
Preview
Text (ABSTRAK)
02 ABSTRAK.pdf

Download (239kB) | Preview
[img] Text (BAB I)
03 BAB 1.pdf
Restricted to Registered users only

Download (177kB)
[img] Text (BAB II)
04 BAB 2.pdf
Restricted to Registered users only

Download (769kB)
[img] Text (BAB III)
05 BAB 3.pdf
Restricted to Registered users only

Download (720kB)
[img] Text (BAB IV)
06 BAB 4.pdf
Restricted to Registered users only

Download (1MB)
[img] Text (BAB V)
07 BAB 5.pdf
Restricted to Registered users only

Download (235kB)
[img] Text (DAFTAR PUSTAKA)
08 DAFTAR PUSTAKA.pdf
Restricted to Registered users only

Download (177kB)
[img] Text (LAMPIRAN)
09 LAMPIRAN.pdf
Restricted to Registered users only

Download (370kB)

Abstract

A pH meter device is a tool that can measure the acidity and alkalinity of a sample, generally the existing pH meter is in the form of a glass electrode which functions as a measure of the number of ions in a sample. In general, the nature of glass that is prone to breaking makes the sensor prone to leaks, this time Ion selective electrode (ISE) is based on polypyrrole and cloth/carbon electrodes. The carbon coating itself uses the screenprinting method, with paraffin wax as a membrane with various thicknesses, the optimum sensor thickness value is obtained at 8mm. then the fabric/carbon sensor is tested with parameters of variations in polypyrrole coating time and the applied voltage during coating. The optimal value was obtained at the voltage variation at 4.5 Volts using the electrodeposition method with a coating time of 30 seconds, in the pH response test using the potentiometry method the slope value was 54.173 mV/decade in the pH response test using pH buffers 4, 7, and 10 where the value is in accordance with IUPAC standards. In the selectivity test results with a thickness variation of 8mm, the results obtained for interfering ions using Ba2+ with a value of

Item Type: Thesis (S1)
Call Number CD: FT/MSN. 23 016
NIM/NIDN Creators: 41318010018
Uncontrolled Keywords: Ion selective electrode (ISE), kain katun rayon, Potensiometri, pasta karbon, Polypyrrole
Subjects: 600 Technology/Teknologi > 620 Engineering and Applied Operations/Ilmu Teknik dan operasi Terapan
600 Technology/Teknologi > 620 Engineering and Applied Operations/Ilmu Teknik dan operasi Terapan > 621 Applied Physics/Fisika terapan
600 Technology/Teknologi > 620 Engineering and Applied Operations/Ilmu Teknik dan operasi Terapan > 621 Applied Physics/Fisika terapan > 621.8 Machine Engineering, Machinery/Teknik Mesin
Divisions: Fakultas Teknik > Teknik Mesin
Depositing User: ADELINA HASNA SETIAWATI
Date Deposited: 01 Apr 2023 07:08
Last Modified: 01 Apr 2023 07:08
URI: http://repository.mercubuana.ac.id/id/eprint/75854

Actions (login required)

View Item View Item